Gauge Repeatability and Bias Check (Type 1 Study: Cg, Cgk, Bias t-Test)
From repeated readings of a reference master, is this gauge repeatable and accurate enough for the tolerance, and is its bias statistically significant?
Type 1 gauge study from repeated readings of one reference master by one operator: the tolerance, bias (mean minus reference), the potential capability Cg from a share of the tolerance against a spread of standard deviations, the capability with bias Cgk, the percentage of tolerance taken by repeatability and by repeatability with bias, a two-sided t-test of the bias (confidence interval of the mean, as the two-sample t-test tool’s t distribution), and the gauge resolution as a share of the tolerance, against your minimum capability and largest resolution share.
Open the calculatorRuns in your browser. No account needed.
How this tool works
Inputs, units and where to find them
| Input | Unit | Typical source |
|---|---|---|
| Number of repeated readings | readings | Number of readings of the master (studies commonly use 25 to 50). |
| Mean of the readings | Average of the repeated readings, in the unit of the tolerance. | |
| Standard deviation of the readings | Sample standard deviation of the readings. | |
| Reference value of the master | Calibrated value of the reference master, from its certificate. | |
| Lower specification limit | Lower limit of the characteristic’s tolerance. | |
| Upper specification limit | Upper limit of the characteristic’s tolerance. | |
| Gauge resolution (smallest step) | Smallest step the gauge displays. | |
| Acceptance settings | ||
| Share of tolerance used for Cg | % | Share of the tolerance the gauge spread may use (20 % is Minitab’s default). |
| Spread in standard deviations | σ | Number of standard deviations taken as the gauge spread (6 covers about 99.73 %). |
| Minimum Cg and Cgk | Minimum Cg and Cgk your quality system requires. | |
| Largest resolution as a share of tolerance | % | Largest gauge resolution allowed as a share of the tolerance, from your quality system. |
| Significance level for the bias test | choice | Chance of calling a bias significant when there is none. |
What each input means
Open an input to see what it is, its units, the worked example and the common mistakes, with a small diagram.
What this means: Number of repeated readings
- What it is
- The number of times the reference master was measured in the study, all by the same operator under the same conditions.
- Units and format
- Enter a number in readings: readings. The form accepts 3 readings to 1000 readings. Use a whole number. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 25 readings as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- More readings make the bias test more sensitive; Cg and Cgk do not change.
- Get it right
- Remove and replace the master between readings so the study includes handling, not just repeated display values.
What this means: Mean of the readings
- What it is
- The average of all the repeated readings of the master, in the same unit as the tolerance, compared with the master’s reference value to find the bias.
- Units and format
- Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 10.0002 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- Moving it away from the reference value raises the bias, lowering Cgk.
- Get it right
- Keep every reading in the average; discarding readings that look odd hides real gauge problems.
What this means: Standard deviation of the readings
- What it is
- The sample standard deviation of the repeated readings, the scatter of the gauge when measuring the same master again and again.
- Units and format
- Enter a plain number with no unit. The form accepts 0.000001 or more. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 0.0008 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- Cg and Cgk fall in proportion as the scatter rises.
- Get it right
- Use the sample standard deviation (divide by n − 1) of the readings themselves, not of their ranges.
What this means: Reference value of the master
- What it is
- The calibrated value of the reference master or setting standard, taken from its calibration certificate, which the gauge readings are compared with.
- Units and format
- Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 10 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- The bias is the mean less this value, so a change here moves Cgk and the bias test.
- Get it right
- Use the certificate value, not the nominal size; ideally the master sits near the middle of the tolerance.
What this means: Lower specification limit
- What it is
- The lowest acceptable value of the characteristic the gauge will check, the lower end of the tolerance the gauge spread is compared with.
- Units and format
- Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 9.98 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- A lower limit widens the tolerance, raising Cg and Cgk.
- Get it right
- For a one-sided tolerance with a natural zero, such as flatness, enter 0.
What this means: Upper specification limit
- What it is
- The highest acceptable value of the characteristic the gauge will check, the upper end of the tolerance.
- Units and format
- Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 10.02 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- A higher limit widens the tolerance, raising Cg and Cgk.
- Get it right
- It must be above the lower limit, in the same unit as the readings.
What this means: Gauge resolution (smallest step)
- What it is
- The smallest change in reading the gauge can display, its resolution, compared with the tolerance to judge whether the gauge can see small differences.
- Units and format
- Enter a plain number with no unit. The form accepts 0.000001 or more. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 0.0001 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- Only moves the resolution check; Cg and Cgk do not change.
- Get it right
- Use the display step actually shown, for example 0.001 mm on a micrometer with a 0.001 mm digit.
Acceptance settings
What this means: Share of tolerance used for Cg
- What it is
- The share of the tolerance that the gauge spread may take up, the numerator of Cg; Minitab’s type 1 study uses 20 % by default.
- Units and format
- Enter a number in %: percent: the number is read as a percentage, so type 8 for 8 %. The form accepts 1% to 100%. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 20% as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- Cg and Cgk rise in proportion to it.
- Get it right
- Use the share your quality system or customer specifies; change it only with their agreement.
What this means: Spread in standard deviations
- What it is
- The number of standard deviations of the readings taken as the gauge’s spread, the denominator of Cg; six standard deviations cover about 99.73 % of normal readings.
- Units and format
- Enter a number in σ: σ. The form accepts 1 σ to 12 σ. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 6 σ as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- Cg and Cgk fall in proportion as the spread multiplier rises.
- Get it right
- Some systems use 4 or 5.15; use the value your quality system specifies.
What this means: Minimum Cg and Cgk
- What it is
- The smallest Cg and Cgk your quality system accepts for a capable gauge, the pass line for both capability indices.
- Units and format
- Enter a plain number with no unit. The form accepts 0.1 to 10. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 1.33 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- Only moves the pass or fail line for Cg and Cgk.
- Get it right
- Take it from the customer or your quality manual; this tool does not set a value.
What this means: Largest resolution as a share of tolerance
- What it is
- The largest gauge resolution you accept as a share of the tolerance, the pass line for the resolution check.
- Units and format
- Enter a number in %: percent: the number is read as a percentage, so type 8 for 8 %. The form accepts 0.01% to 100%. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 5% as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- Only moves the pass or fail line for resolution.
- Get it right
- Take it from your quality system; a gauge that cannot resolve small steps hides real variation.
What this means: Significance level for the bias test
- What it is
- The risk accepted of calling the gauge biased when its true bias is zero, used as the threshold for the p-value of the bias t-test.
- Units and format
- Choose one of 3 options from the list: 0.10; 0.05; 0.01.
- Example
- The form opens with “0.05” selected as a worked example. Choose the option that matches your case.
- Effect on the result
- Only changes whether the bias is reported as significant.
- Get it right
- 0.05 is usual; a smaller value demands stronger evidence before calling the bias real.
Method and formulas
- Tolerance T = USL − LSL; bias = x̄ − x_ref.
- Cg = (K × T) ÷ (L s); Cgk = (K × T ÷ 2 − |bias|) ÷ (L s ÷ 2); share of tolerance for repeatability = L s ÷ T and with bias = K ÷ Cgk.
- Bias test: t = bias ÷ (s ÷ √n) with n − 1 degrees of freedom; p from the t distribution (two-sided); significant when p < α.
- Example: Cg = 1.6667 and Cgk = 1.5833; the bias of 0.0002 has p = 0.22335.
Limitations
- One master at one point of the range, one operator: a type 1 study shows repeatability and bias only, not reproducibility between operators or linearity across the range.
- Readings assumed independent and roughly normal; drift during the study inflates the standard deviation.
- Example values are illustrations. Replace them with values for your actual task.
Typical uses
- Accept a new gauge or CMM program for a characteristic
- Check a gauge after repair or recalibration
- Decide whether bias needs correcting
Who it is for
- Quality engineers
- Metrology technicians
- Production supervisors
What you get
- Tolerance, bias and t-test
- Cg and Cgk
- Share of tolerance used
- Resolution share and the decision
Interpreting the result
- A gauge with good Cg but poor Cgk is repeatable but offset: correcting the bias (re-mastering or an offset) can be enough.
References
- Machinery’s Handbook, 29th edition (Industrial Press, 2012), printed pp. 128–129. Topic reference; no source tables or recommended settings are reproduced.
- Minitab, Methods and formulas for Type 1 Gage Study — https://support.minitab.com/en-us/minitab/help-and-how-to/quality-and-process-improvement/measurement-system-analysis/how-to/gage-study/type-1-gage-study/methods-and-formulas/methods-and-formulas/
- VDA Volume 5 (Verband der Automobilindustrie), capability of measurement processes: the source of the type 1 study; consult the current edition.
Learn the Concepts Behind This Tool
These KEVOS® Library readings explain the method this tool uses, with worked examples.
Explains the concept · Library collection
Measurement system analysis: making sure your gauges can tell good parts from bad
What it covers: Every accept, reject and adjust decision trusts a measurement. How to check bias, repeatability and reproducibility, run a gauge R&R study, judge the result and fix a weak system.
Method explained here · KEVOS® original book
How good is the gauge?
In Accept or Reject by the Numbers
What it covers: Judge a gauge three ways: a Type 1 study of repeatability and bias with the KEVOS® gauge check, the test uncertainty ratio, and the money a better gauge would save in wrong decisions.
Method explained here · KEVOS® original book
Measuring conductivity reliably
In Eddy-Current Inspection by the Numbers
What it covers: Edge distance, curvature, lift-off and probe condition in conductivity readings, a Type 1 gauge study on a reference standard, and an uncertainty budget for the acceptance band.
Method explained here · KEVOS® original book
Accuracy classes and whether a scale is good enough
In Strain Gauges and Load Cells by the Numbers
What it covers: How to read a load cell's OIML accuracy class and error limits, when trade rules apply, and how a gauge study and error budget show whether a scale can weigh to a tolerance.
KEVOS® tools are screening and decision-support utilities. Verify important financial, engineering, safety, legal or operational decisions using appropriate professional standards and your organisation’s approved data.
Open the Gauge Repeatability and Bias Check (Type 1 Study: Cg, Cgk, Bias t-Test) calculator