Gauge Repeatability and Bias Check (Type 1 Study: Cg, Cgk, Bias t-Test)

From repeated readings of a reference master, is this gauge repeatable and accurate enough for the tolerance, and is its bias statistically significant?

Type 1 gauge study from repeated readings of one reference master by one operator: the tolerance, bias (mean minus reference), the potential capability Cg from a share of the tolerance against a spread of standard deviations, the capability with bias Cgk, the percentage of tolerance taken by repeatability and by repeatability with bias, a two-sided t-test of the bias (confidence interval of the mean, as the two-sample t-test tool’s t distribution), and the gauge resolution as a share of the tolerance, against your minimum capability and largest resolution share.

Open the calculatorRuns in your browser. No account needed.

How this tool works

Inputs, units and where to find them

InputUnitTypical source
Number of repeated readingsreadingsNumber of readings of the master (studies commonly use 25 to 50).
Mean of the readingsAverage of the repeated readings, in the unit of the tolerance.
Standard deviation of the readingsSample standard deviation of the readings.
Reference value of the masterCalibrated value of the reference master, from its certificate.
Lower specification limitLower limit of the characteristic’s tolerance.
Upper specification limitUpper limit of the characteristic’s tolerance.
Gauge resolution (smallest step)Smallest step the gauge displays.
Acceptance settings
Share of tolerance used for Cg%Share of the tolerance the gauge spread may use (20 % is Minitab’s default).
Spread in standard deviationsσNumber of standard deviations taken as the gauge spread (6 covers about 99.73 %).
Minimum Cg and CgkMinimum Cg and Cgk your quality system requires.
Largest resolution as a share of tolerance%Largest gauge resolution allowed as a share of the tolerance, from your quality system.
Significance level for the bias testchoiceChance of calling a bias significant when there is none.

What each input means

Open an input to see what it is, its units, the worked example and the common mistakes, with a small diagram.

What this means: Number of repeated readings
Illustration: Number of repeated readingsA row of identical items being counted; Number of repeated readings is how many there are.nExample: 25 readingsNumber of repeated readings
What it is
The number of times the reference master was measured in the study, all by the same operator under the same conditions.
Units and format
Enter a number in readings: readings. The form accepts 3 readings to 1000 readings. Use a whole number. Required: the tool does not calculate until every input is valid.
Example
The form opens with 25 readings as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
More readings make the bias test more sensitive; Cg and Cgk do not change.
Get it right
Remove and replace the master between readings so the study includes handling, not just repeated display values.
What this means: Mean of the readings
Illustration: Mean of the readingsA number line with a marker at the value of Mean of the readings, between its lowest and highest accepted values.−10⁹10⁹x̄Example: 10.0002Mean of the readings
What it is
The average of all the repeated readings of the master, in the same unit as the tolerance, compared with the master’s reference value to find the bias.
Units and format
Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
Example
The form opens with 10.0002 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Moving it away from the reference value raises the bias, lowering Cgk.
Get it right
Keep every reading in the average; discarding readings that look odd hides real gauge problems.
What this means: Standard deviation of the readings
Illustration: Standard deviation of the readingsA nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Standard deviation of the readings.upperlowersExample: 0.0008Standard deviation of the readings
What it is
The sample standard deviation of the repeated readings, the scatter of the gauge when measuring the same master again and again.
Units and format
Enter a plain number with no unit. The form accepts 0.000001 or more. Required: the tool does not calculate until every input is valid.
Example
The form opens with 0.0008 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Cg and Cgk fall in proportion as the scatter rises.
Get it right
Use the sample standard deviation (divide by n − 1) of the readings themselves, not of their ranges.
What this means: Reference value of the master
Illustration: Reference value of the masterA number line with a marker at the value of Reference value of the master, between its lowest and highest accepted values.−10⁹10⁹x_refExample: 10Reference value of the master
What it is
The calibrated value of the reference master or setting standard, taken from its calibration certificate, which the gauge readings are compared with.
Units and format
Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
Example
The form opens with 10 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
The bias is the mean less this value, so a change here moves Cgk and the bias test.
Get it right
Use the certificate value, not the nominal size; ideally the master sits near the middle of the tolerance.
What this means: Lower specification limit
Illustration: Lower specification limitA nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Lower specification limit.upperlowerLSLExample: 9.98Lower specification limit
What it is
The lowest acceptable value of the characteristic the gauge will check, the lower end of the tolerance the gauge spread is compared with.
Units and format
Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
Example
The form opens with 9.98 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
A lower limit widens the tolerance, raising Cg and Cgk.
Get it right
For a one-sided tolerance with a natural zero, such as flatness, enter 0.
What this means: Upper specification limit
Illustration: Upper specification limitA nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Upper specification limit.upperlowerUSLExample: 10.02Upper specification limit
What it is
The highest acceptable value of the characteristic the gauge will check, the upper end of the tolerance.
Units and format
Enter a plain number with no unit. Negative values are accepted. Required: the tool does not calculate until every input is valid.
Example
The form opens with 10.02 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
A higher limit widens the tolerance, raising Cg and Cgk.
Get it right
It must be above the lower limit, in the same unit as the readings.
What this means: Gauge resolution (smallest step)
Illustration: Gauge resolution (smallest step)A nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Gauge resolution (smallest step).upperlower±Example: 0.0001Gauge resolution (smallest step)
What it is
The smallest change in reading the gauge can display, its resolution, compared with the tolerance to judge whether the gauge can see small differences.
Units and format
Enter a plain number with no unit. The form accepts 0.000001 or more. Required: the tool does not calculate until every input is valid.
Example
The form opens with 0.0001 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Only moves the resolution check; Cg and Cgk do not change.
Get it right
Use the display step actually shown, for example 0.001 mm on a micrometer with a 0.001 mm digit.

Acceptance settings

What this means: Share of tolerance used for Cg
Illustration: Share of tolerance used for CgA bar of 100 parts with the share marked; Share of tolerance used for Cg is that share out of 100. The bar follows the value you type.0 %100 %KExample: 20%Share of tolerance used for Cg
What it is
The share of the tolerance that the gauge spread may take up, the numerator of Cg; Minitab’s type 1 study uses 20 % by default.
Units and format
Enter a number in %: percent: the number is read as a percentage, so type 8 for 8 %. The form accepts 1% to 100%. Required: the tool does not calculate until every input is valid.
Example
The form opens with 20% as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Cg and Cgk rise in proportion to it.
Get it right
Use the share your quality system or customer specifies; change it only with their agreement.
What this means: Spread in standard deviations
Illustration: Spread in standard deviationsTwo bars: a base value and the same value multiplied by Spread in standard deviations, the factor.base valuewith factorLExample: 6 σSpread in standard deviations
What it is
The number of standard deviations of the readings taken as the gauge’s spread, the denominator of Cg; six standard deviations cover about 99.73 % of normal readings.
Units and format
Enter a number in σ: σ. The form accepts 1 σ to 12 σ. Required: the tool does not calculate until every input is valid.
Example
The form opens with 6 σ as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Cg and Cgk fall in proportion as the spread multiplier rises.
Get it right
Some systems use 4 or 5.15; use the value your quality system specifies.
What this means: Minimum Cg and Cgk
Illustration: Minimum Cg and CgkTwo bars: a base value and the same value multiplied by Minimum Cg and Cgk, the factor.base valuewith factor×Example: 1.33Minimum Cg and Cgk
What it is
The smallest Cg and Cgk your quality system accepts for a capable gauge, the pass line for both capability indices.
Units and format
Enter a plain number with no unit. The form accepts 0.1 to 10. Required: the tool does not calculate until every input is valid.
Example
The form opens with 1.33 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Only moves the pass or fail line for Cg and Cgk.
Get it right
Take it from the customer or your quality manual; this tool does not set a value.
What this means: Largest resolution as a share of tolerance
Illustration: Largest resolution as a share of toleranceA bar of 100 parts with the share marked; Largest resolution as a share of tolerance is that share out of 100. The bar follows the value you type.0 %100 %Example: 5%Largest resolution as a share of tolerance
What it is
The largest gauge resolution you accept as a share of the tolerance, the pass line for the resolution check.
Units and format
Enter a number in %: percent: the number is read as a percentage, so type 8 for 8 %. The form accepts 0.01% to 100%. Required: the tool does not calculate until every input is valid.
Example
The form opens with 5% as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Only moves the pass or fail line for resolution.
Get it right
Take it from your quality system; a gauge that cannot resolve small steps hides real variation.
What this means: Significance level for the bias test
Illustration: Significance level for the bias testThe options for Significance level for the bias test, with the one currently selected highlighted.0.100.050.01Example: 0.05Significance level for the bias test
What it is
The risk accepted of calling the gauge biased when its true bias is zero, used as the threshold for the p-value of the bias t-test.
Units and format
Choose one of 3 options from the list: 0.10; 0.05; 0.01.
Example
The form opens with “0.05” selected as a worked example. Choose the option that matches your case.
Effect on the result
Only changes whether the bias is reported as significant.
Get it right
0.05 is usual; a smaller value demands stronger evidence before calling the bias real.

Method and formulas

  1. Tolerance T = USL − LSL; bias = x̄ − x_ref.
  2. Cg = (K × T) ÷ (L s); Cgk = (K × T ÷ 2 − |bias|) ÷ (L s ÷ 2); share of tolerance for repeatability = L s ÷ T and with bias = K ÷ Cgk.
  3. Bias test: t = bias ÷ (s ÷ √n) with n − 1 degrees of freedom; p from the t distribution (two-sided); significant when p < α.
  4. Example: Cg = 1.6667 and Cgk = 1.5833; the bias of 0.0002 has p = 0.22335.

Limitations

  • One master at one point of the range, one operator: a type 1 study shows repeatability and bias only, not reproducibility between operators or linearity across the range.
  • Readings assumed independent and roughly normal; drift during the study inflates the standard deviation.
  • Example values are illustrations. Replace them with values for your actual task.

Typical uses

  • Accept a new gauge or CMM program for a characteristic
  • Check a gauge after repair or recalibration
  • Decide whether bias needs correcting

Who it is for

  • Quality engineers
  • Metrology technicians
  • Production supervisors

What you get

  • Tolerance, bias and t-test
  • Cg and Cgk
  • Share of tolerance used
  • Resolution share and the decision

Interpreting the result

  • A gauge with good Cg but poor Cgk is repeatable but offset: correcting the bias (re-mastering or an offset) can be enough.

References

  • Machinery’s Handbook, 29th edition (Industrial Press, 2012), printed pp. 128–129. Topic reference; no source tables or recommended settings are reproduced.
  • Minitab, Methods and formulas for Type 1 Gage Study — https://support.minitab.com/en-us/minitab/help-and-how-to/quality-and-process-improvement/measurement-system-analysis/how-to/gage-study/type-1-gage-study/methods-and-formulas/methods-and-formulas/
  • VDA Volume 5 (Verband der Automobilindustrie), capability of measurement processes: the source of the type 1 study; consult the current edition.