KEVOS® MAKEOS · Quality
Instrument Calibration Check (Errors, Accuracy Class, Interpolated Correction)
From my calibration readings against a reference, is this gauge within its accuracy class, and what is the corrected value of a reading taken in service?
Calibration check of a pressure gauge, thermometer or other instrument at five points across its range: the error at each point and as a percentage of span against the accuracy class, the worst point, and the corrected (true) value of an in-service reading by linear interpolation between the two calibration points that bracket it (as the linear interpolation tool).
Open the calculatorRuns in your browser. No account needed.
How this tool works
Inputs, units and where to find them
| Input | Unit | Typical source |
|---|---|---|
| Range low end | Lowest value of the instrument range. | |
| Range high end | Highest value of the instrument range. | |
| Accuracy class (% of span) | % | Permitted error as a percentage of span. |
| Reference values and readings | ||
| Reference 1 | Reference (true) value at calibration point 1. | |
| Reading 1 | Instrument reading at point 1. | |
| Reference 2 | Reference value at point 2. | |
| Reading 2 | Reading at point 2. | |
| Reference 3 | Reference value at point 3. | |
| Reading 3 | Reading at point 3. | |
| Reference 4 | Reference value at point 4. | |
| Reading 4 | Reading at point 4. | |
| Reference 5 | Reference value at point 5. | |
| Reading 5 | Reading at point 5. | |
| In service | ||
| Reading taken in service | A reading taken in use that you want corrected. | |
Method and formulas
- Error = reading − reference; % of span = error ÷ (high − low) × 100.
- Within class if every |error| ≤ class × span ÷ 100.
- Corrected value: find the true value for the reading by linear interpolation between the bracketing calibration points (as the linear interpolation tool, finding x for a given y).
- Example: the worst error is 0.4% of span against 1%; a reading of 6 corresponds to 5.9677.
Limitations
- Rising readings only; hysteresis needs falling readings as well.
- Readings must increase with the reference for interpolation.
- Example values are illustrations. Replace them with values for your actual task.
Typical uses
- Record and judge a gauge calibration
- Correct readings from a gauge with a known error
- Decide whether to adjust or replace an instrument
Who it is for
- Instrument technicians
- Quality technicians
- Maintenance planners
What you get
- Error at each point and as % of span
- Worst error against the class
- Corrected value of the service reading
Interpreting the result
- If the errors are all one way, a zero or span adjustment usually brings the instrument back into class.
References
- Machinery’s Handbook, 29th edition (Industrial Press, 2012), printed pp. 121, 709–735. Topic reference; no source tables or recommended settings are reproduced.
WORKOS tools are screening and decision-support utilities. Verify important financial, engineering, safety, legal or operational decisions using appropriate professional standards and your organisation’s approved data.
Open the Instrument Calibration Check (Errors, Accuracy Class, Interpolated Correction) calculator